Description
Premium Getinge HS-200 encapsulated reverse osmosis filter (61301605455). Includes replacement flow restrictor. Made in the USA in ISO 9001:2008 Certified Facilities. Replace your old Getinge RO Membrane with this Superior Premium Replacement.
Membranes required for Getinge / Watts HS-Series water systems:
- HS-200 requires three (3) membrane filters
Specifications
Recommended Operation Conditions:
• Maximum operation pressure: 125 psi (0.86 MPa)
• Maximum feed flow rate: 2 gpm
• Maximum operating temperature: 113° F (45° C)
• Maximum feed water turbidity: 1 NTU
• Maximum feed water SDI (15min): 5
• Chlorine tolerance: 0
• Feedwater pH range: 2-11
• Minimum brine flow to permeate flow ratio: 4:1
Performance specifications based on 500 ppm tap water, 50psi (0.35 MPa) applied pressure, 77°F (25°C) feed water temperature, feed water pH 7-8 and 15% recovery. Element permeate flow may vary ± 15%.